Objectives: The incidence of femoral neck fractures in younger-age people increases due to high-energy trauma. To avoid complications, stable internal fixation is essential. The most commonly used implants for vertically oriented femoral neck fractures are the cannulated and dynamic hip screws. In our study, we compared differently oriented cannulated screw configurations and a standard DHS for fixation of femoral neck fracture.Materials and Methods: A finite element-based collum femoris fracture was created and was fixed using four differently oriented triangular screw configurations and a dynamic hip screw. The loads were applied using a commercially available software package.Results: Centrally oriented cannulated screw configuration had the most compression and compression stress on the fracture side, and it had the highest stress values on the implant.Conclusion: We recommend the use of centrally oriented triangular cannulated screws for femoral neck fracture fixation. With this construct, more compression and compression pressure can be obtained, which can result in the early healing of the fracture. However, it must be kept in mind that this construct exerts more stress on the implant, which could be a reason for implant failure

Keywords

References

  1. Pauyo T, Drager J, Albers A, Harvey EJ. Management of femoral neck fractures in the young patient: A critical analysis review. World journal of orthopedics. 2014;5:204-17. [CrossRef]
  2. Johansson T, Jacobsson SA, Ivarsson I, Knutsson A, Wahlstrom O. Internal fixation versus total hip arthroplasty in the treatment of displaced femoral neck fractures: a prospective randomized study of 100 hips. Acta orthopaedica Scandinavica. 2000;71:597-602. [CrossRef]
  3. Luttrell K, Beltran M, Collinge CA. Preoperative decision making in the treatment of high-angle “vertical” femoral neck fractures in young adult patients. An expert opinion survey of the Orthopaedic Trauma Association’s (OTA) membership. Journal of orthopaedic trauma. 2014;28:e221-5. [CrossRef]
  4. Dedrick DK, Mackenzie JR, Burney RE. Complications of femoral neck fracture in young adults. The Journal of trauma. 1986;26:932-7.
  5. Zhang Y, Tian L, Yan Y, Sang H, Ma Z, Jie Q, et al. Biomechanical evaluation of the expansive cannulated screw for fixation of femoral neck fractures. Injury. 2011;42:1372-6. [CrossRef]
  6. Kunapuli SC, Schramski MJ, Lee AS, Popovich JM, Jr., Cholewicki J, Reeves NP, et al. Biomechanical Analysis of Augmented Plate Fixation for the treatment of Vertical Shear Femoral Neck Fractures. Journal of orthopaedic trauma. 2015;29:144-50. [CrossRef]
  7. Rupprecht M, Grossterlinden L, Sellenschloh K, Hoffmann M, Puschel K, Morlock M, et al. Internal fixation of femoral neck fractures with posterior comminution: a biomechanical comparison of DHS(R) and Intertan nail(R). International orthopaedics. 2011;35:1695-701. [CrossRef]
  8. Rupprecht M, Grossterlinden L, Ruecker AH, de Oliveira AN, Sellenschloh K, Nuchtern J, et al. A comparative biomechanical analysis of fixation devices for unstable femoral neck fractures: the Intertan versus cannulated screws or a dynamic hip screw. The Journal of trauma. 2011;71:625-34. [CrossRef]

Details

Primary Language

English

Subjects

-

Journal Section

-

Authors

Okan Özkunt

Kerim Sarıyılmaz

Turgut Akgül

Gökhan Polat

Fatih Dikici

Publication Date

September 1, 2018

Submission Date

-

Acceptance Date

-

Published in Issue

Year 2018 Number: 3

EndNote
Özkunt O, Sarıyılmaz K, Akgül T, Polat G, Dikici F (September 1, 2018) Does The Screw Orientation Effects The Stability of Femoral Neck Fracture?: a Finite Element Analysis. Acıbadem Üniversitesi Sağlık Bilimleri Dergisi 3 243–247.